ARTHREX, INC. ACL TIGHTROPE RT W/ DEPLOYING SUTURE; SMOOTH FIXATION PIN Back to Search Results
Model Number ACL TIGHTROPE RT W/ DEPLOYING SUTURE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint is confirmed.One unpackaged ar-1588rt-j was received for investigation.Functional testing was not able to be performed due to the damage to the device.Visual evaluation found suture was torn.No sharp edges were observed on the button.The most likely cause is attributed to user error due to excessive force.Per dfu-0147.Precautions.Acl/pcl/ btb/rt/abs tightrope only: excessive force on the shortening suture strands may break the strands and impair ability to fully seat the implant.No additional force on the shortening suture strands is required when the graft/wedge construct reaches the desired position in the femoral socket and the graft stability is verified by pulling distally on the graft.Refer to investigation photos.
 
Event or Problem Description
It was reported that during a knee surgery the white suture of the device tore.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.(b)(6) 2023 update dw further information were provided that the torn pieces were retrieved out of the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE RT W/ DEPLOYING SUTURE
Common Device Name
SMOOTH FIXATION PIN
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19574706
Report Number1220246-2024-06267
Device Sequence Number19819468
Product Code MBI
UDI-Device Identifier00888867095533
UDI-Public00888867095533
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K231857
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACL TIGHTROPE RT W/ DEPLOYING SUTURE
Device Catalogue NumberAR-1588RT-J
Device Lot Number15111317
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2023
Initial Report FDA Received Date06/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-