| Catalog Number |
CDS0601-NTR |
| Medical Device Problem Code |
Incomplete Coaptation (2507)
|
| Health Effect - Clinical Code |
Mitral Valve Insufficiency/ Regurgitation (4451)
|
| Date of Event |
05/29/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The clip remains in the patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2020, a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with a grade of 4.One clip was implanted, reducing mr to a grade of <1.On (b)(6) 2024, echocardiogram showed the implanted clip had detached from posterior leaflet and remained attached to anterior leaflet (single leaflet device attachment/slda), causing mr to increase to a grade of 3-4.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, a cause for the slda could not be conclusively determined.The reported mr is a cascading event of the slda.The reported patient effect of mitral regurgitation, as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.The reported serious injury/ illness/ impairment was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|