ABBOTT MEDICAL MITRACLIP; MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0601-NTR
Medical Device Problem Code Incomplete Coaptation (2507)
Health Effect - Clinical Code Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 05/29/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The clip remains in the patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2020, a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with a grade of 4.One clip was implanted, reducing mr to a grade of <1.On (b)(6) 2024, echocardiogram showed the implanted clip had detached from posterior leaflet and remained attached to anterior leaflet (single leaflet device attachment/slda), causing mr to increase to a grade of 3-4.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, a cause for the slda could not be conclusively determined.The reported mr is a cascading event of the slda.The reported patient effect of mitral regurgitation, as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.The reported serious injury/ illness/ impairment was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19575047
Report Number2135147-2024-02912
Device Sequence Number18245240
Product Code NKM
UDI-Device Identifier08717648226342
UDI-Public(01)08717648226342(17)201213(10)91213U102
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2020
Device Catalogue NumberCDS0601-NTR
Device Lot Number91213U102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/29/2024
Supplement Date Received by Manufacturer07/03/2024
Initial Report FDA Received Date06/20/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/14/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient SexUnknown
-
-