MAUDE Adverse Event Report: BAUSCH + LOMB ENVISTA SIMPLIFEYE INSERTER; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB ENVISTA SIMPLIFEYE INSERTER; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Model Number 21987
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was discarded and therefore not available for evaluation.A review of device history record (dhr) did not find any non-conformities or anomalies related to the reported event.The lot history, trend analysis, risk analysis, and directions for use are considered acceptable with the product performing within anticipated rates.Based on the available information, the root cause of this event could not be determined.; however, user related factors (such as loading or handling techniques) and/or procedural factors (such as lens and inserter interaction) might have contributed to the event.
 
Event or Problem Description
It was that reported during implantation of an intraocular lens (iol) into the patients eye the surgeon noticed the optic of the lens was scratched.The incision was enlarged to allow for removal of the iol and a successful intraoperative lens exchange was performed using a backup iol.One suture was used to close the wound.Patient outcome is good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVISTA SIMPLIFEYE INSERTER
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
MEDICEL AG
dornierstrasse 11
altenrheim CH-94 23
SZ   CH-9423
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key19578841
Report Number0001313525-2024-00115
Device Sequence Number8525467
Product Code MSS
UDI-Device Identifier00757770547449
UDI-Public(01)00757770547449(17)260831(10)2304A1
Combination Product (Y/N)N
PMA/510(K) Number
K192005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model Number21987
Device Lot Number2304A1
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2024
Initial Report FDA Received Date06/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENVISTA INTRAOCULAR LENS.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-