| Catalog Number |
81660100S |
| Medical Device Problem Code |
Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that: after lag screw inserted, compression applied, set screw tightened and slightly loosen, when u-blade was inserted, the femoral head was pushed and the fracture separated.The surgery was finished with pulling the femoral head back somewhat.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was implanted and no other evidence was provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.After lag screw inserted, compression applied, set screw tightened and slightly loosen, when u-blade was inserted, the femoral head was pushed and the fracture separated.The surgery was finished with pulling the femoral head back somewhat.More detailed information about the complaint event, e.G.X-ray in m-l- view, must be available in order to determine the root cause of the complaint event.At the end of the surgery, it is the responsibility of the surgeon if he leaves the patient as is.Based on investigation, the root cause was attributed to a user related issue.With available information a product deficiency was not verified.In case we receive further substantial information we reserve the right re-open the case and to re-assess the root cause.
|
| |
|
Event or Problem Description
|
|
It was reported that: after lag screw inserted, compression applied, set screw tightened and slightly loosen, when u-blade was inserted, the femoral head was pushed and the fracture separated.The surgery was finished with pulling the femoral head back somewhat.
|
| |
|