| Catalog Number |
02K43-20 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed that quality control were out of range high for all three levels while using architect stat myoglobin reagent list 02k43-20, lot 60104un23 on the architect i1000 processing module.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation into this issue found that during manufacturing, the microparticle concentrate used in reagent lot 60104un23 was not properly mixed.As a result, higher percentage of microparticles was added to reagent lot number 60104un23.A product recall letter was issued on 21feb2024 to all customers who have received shipments of architect stat myoglobin reagent lot number 60104un23.The product recall letter notifies the customer of the issue and the potential impact to patient results with use of lot 60104un23.The product recall letter instructs customers to discontinue use of and destroy any remaining inventory of lot 60104un23 according to their laboratory procedures and immediately contact customer support to order a replacement product.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up being submitted to update the section h6: component code from g03001 analyzer to g01003 device ingredient or reagent.
|
| |
|
Event or Problem Description
|
|
The customer observed that quality control were out of range high for all three levels while using architect stat myoglobin reagent list 02k43-20, lot 60104un23 on the architect i1000 processing module.There was no reported impact to patient management.
|
| |
|