ZIMMER GMBH PROXIMAL HUMERUS, LEFT, LONG, ÿ 8.5X220MM; TRAUMA, INSTRUMENTS Back to Search Results
Model Number N/A
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 05/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.Review of complaint histories identified additional similar complaints for the reported products and additional similar complaints for the reported part and lot combinations.Based on the available information, the reported event cannot be confirmed.It can be assumed that multiple contributing factors, related to either patient condition and behavior or implantation procedure, contributed to the migration of the screws.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism of the affixus nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the affixus® natural nail® proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).D10: 47-2486-038-40 ann blunt tip screw 4x38mm 3185170.47-2486-040-40 ann blunt tip screw 4x40mm 3181917.47-2486-042-40 ann blunt tip screw 4x42mm 3181919.47-2486-128-40 ann cort bone screw 4 x 28mm 3191519.47-2486-130-40 ann cort bone screw 4 x 30mm 3177549.47-2488-010-00 affixus ph nl cap 0mm 3189257.G2: report source japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported patient underwent revision approximately 4 weeks from the initial surgery, surgeon found one of the proximal screws was backed out from the proper position.The patient complained of pain and a revision was performed to remove the migrated screw.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROXIMAL HUMERUS, LEFT, LONG, ÿ 8.5X220MM
Common Device Name
TRAUMA, INSTRUMENTS
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key19579343
Report Number0009613350-2024-00246
Device Sequence Number18260223
Product Code HSB
UDI-Device Identifier00889024505933
UDI-Public(01)00889024505933(17)331116(10)3177532
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200814
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number47249622108
Device Lot Number3177532
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2024
Supplement Date Received by Manufacturer07/31/2024
Initial Report FDA Received Date06/20/2024
Supplement Report FDA Received Date08/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SE H11
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-