| Model Number |
560BCS1 |
| Medical Device Problem Codes |
Battery Problem (2885); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: during preventive maintenance by service technician this bio-console 560 instrument's batteries did not discharge properly and blind connector was broken.The issue was resolved by replacing the pcb assy, the gasket, the battery and calibration seal.Preventive maintenance was completed per specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
During preventive maintenance by a service technician this bio-console 560 instrument's batteries did not discharge properly and the blind connector was broken.This was detected during service so there was no patient involvement, so no adverse effect occurred.Medtronic received additional information that the battery did not reach its expected shelf life.The battery had been installed in the instrument since february 18th, 2022.The lot number of the battery removed from the instrument was #13458945.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|