MAUDE Adverse Event Report: ARTHREX, INC. DW OUTFLOW CASSETTE/TUBE SET; ARTHROSCOPE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DW OUTFLOW CASSETTE/TUBE SET; ARTHROSCOPE AND ACCESSORIES Back to Search Results
Model Number DW OUTFLOW CASSETTE/TUBE SET
Medical Device Problem Code Increase in Pressure (1491)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that an external medical technician consultant has found a problem with increasing pressure and the error message "pressure to high" was displayed.There was no harm for patient, operator or third party reported.This was noticed during the testing of the external technician.There was no patient involvement.No further information received.Update dw 27-may-2024: further information were provided that the issue occurred during an arthroscopic hip surgery on (b)(6) 2024.It was further reported that the surgery could be finished successfully with a change of the pump and tubes.No additional surgery is planned however the patient experienced severe pain for a period of three days following her surgical procedure.
 
Additional Manufacturer Narrative
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device.The complaint allegation is not confirmed due to our inability to do functional testing in the lab.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DW OUTFLOW CASSETTE/TUBE SET
Common Device Name
ARTHROSCOPE AND ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19581730
Report Number1220246-2024-06316
Device Sequence Number3266488
Product Code HRX
UDI-Device Identifier00888867039322
UDI-Public00888867039322
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K083707
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDW OUTFLOW CASSETTE/TUBE SET
Device Catalogue NumberAR-6430
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2024
Supplement Date Received by Manufacturer08/21/2024
Initial Report FDA Received Date06/21/2024
Supplement Report FDA Received Date09/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-