| Model Number |
MR850AFU |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The subject mr850 respiratory humidifier has been requested to be returned to fisher & paykel healthcare in new zealand for evaluation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in france reported that a mr850 respiratory humidifier was found with a faulty speaker.Upon device servicing at the regional office in france, it was confirmed that the audible alarm was not functioning.There was no reported patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Correction: section d1: brand name corrected to fisher & paykel healthcare.Section h11: method: the subject pcb of the mr850 respiratory humidifier was returned to fisher & paykel (f&p) healthcare in new zealand for evaluation.Results: testing of the subject pcb confirmed that no sound was generated from the speaker.Additional resistance testing identified an open circuit in the speaker's coil.Conclusion: the cause of the speaker failure was an open circuit in the speaker coil.The mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 is equipped with visual alarm indicators in addition to the audible alarm.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in france reported that a mr850 respiratory humidifier was found with a faulty speaker.Upon device servicing at the regional office in france, it was confirmed that the audible alarm was not functioning.There was no reported patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Correction: section g3: date corrected.Section h11: method: the subject pcb of the mr850 respiratory humidifier was returned to fisher & paykel (f&p) healthcare in new zealand for evaluation.Results: testing of the subject pcb confirmed that no sound was generated from the speaker.Additional resistance testing identified an open circuit in the speaker's coil.Conclusion: the cause of the speaker failure was an open circuit in the speaker coil.The mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 is equipped with visual alarm indicators in addition to the audible alarm.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in france reported that a mr850 respiratory humidifier was found with a faulty speaker.Upon device servicing at the regional office in france, it was confirmed that the audible alarm was not functioning.There was no reported patient involvement.
|
| |
|