• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS FEM TRIAL SZ 5 LT; FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS FEM TRIAL SZ 5 LT; FEMORAL TRIALS Back to Search Results
Catalog Number 254500735
Medical Device Problem Codes Naturally Worn (2988); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the attune femoral trials were scratched.
 
Additional Manufacturer Narrative
Product complaint(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received: a.It was stated that the plastic orange clamping ring is damaged.Please clarify what do you mean by damage? was it broke, cracked, bent or any other deficiency with the device that was not function as intended? it has a chip in it.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : broken attune 2 x femoral trials, patella clamp and femoral introducer.Reported broken by cssd staff.Femoral holder red knob broken off.2 x femoral trials scratched.Patella clamping ring plastic damaged.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual inspection of returned device revealed deep scratches on the top surface.No other defect was found.This type of damage is consistent with the device coming in contact with the saw blade while trailing, care should be taken to avoid saw blade excursion into the femoral trials or implants.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune ps fem trial sz 5 lt would contribute to the complained device issue.¿ based on the investigation findings, the potential cause is traced to unintended use error caused or contributed to event, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM TRIAL SZ 5 LT
Common Device Name
FEMORAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19582037
Report Number1818910-2024-13499
Device Sequence Number14329125
Product Code HWT
UDI-Device Identifier10603295134077
UDI-Public(01)10603295134077(10)MVMAYH710
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500735
Device Lot NumberMVMAYH710
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2024
Supplement Date Received by Manufacturer06/24/2024
08/07/2024
09/06/2024
Initial Report FDA Received Date06/21/2024
Supplement Report FDA Received Date06/26/2024
08/08/2024
09/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-