| Model Number |
MR8-AVA07 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating.No patient impact reported.Repair request escalated to product event due to reason for return.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation determined that the device was overheating and the maximum temperature was measured to be 166f.The likely cause of failure is due to distal bearing is detached.It was also noted that the color ring is scratched/faded and the laser markings are faded.Date of notification: 2024-may-27.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|