MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MR8 ANGLED VAR ATTACHMENT; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MR8 ANGLED VAR ATTACHMENT; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number MR8-AVA07
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Repair request initiated for device with the report of overheating.No patient impact reported.Repair request escalated to product event due to reason for return.
 
Additional Manufacturer Narrative
H3: product analysis: evaluation determined that the device was overheating and the maximum temperature was measured to be 166f.The likely cause of failure is due to distal bearing is detached.It was also noted that the color ring is scratched/faded and the laser markings are faded.Date of notification: 2024-may-27.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MR8 ANGLED VAR ATTACHMENT
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
glen belmer
7000 central avenue ne
ft. worth, TX 76137-3291
6122713209
MDR Report Key19582408
Report Number1625507-2024-00184
Device Sequence Number12610475
Product Code HBB
UDI-Device Identifier00643169912649
UDI-Public00643169912649
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AVA07
Device Catalogue NumberMR8-AVA07
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2024
Initial Report FDA Received Date06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-