It was reported that the balloon of a fogarty catheter was found with pinholes.Catheter was opened and prepped in the sterile filed when holes were noticed.There was no patient involvement.A product evaluation was completed with the returned 120805f.The reported event of balloon issue was confirmed.Balloon latex appeared deteriorated.Multiple cracks and tears were evident on balloon latex.Leakage was observed through the tears on the balloon.Both balloon windings were intact.Balloon latex was released from proximal winding to check balloon edges at the tears.The edges did not appear to match at the tears.No other visible damage was observed from catheter body.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.Based on the available information, there is no evidence that supports or confirms the failure mode is associated to a manufacturing or design defect.A capa was previously initiated and used as reference in the evaluation as it addresses the "balloon - latex deterioration" failure for the embolectomy models with a vascupouch packaging.The capa investigation concluded that the root cause is exposure of latex to ozone, which can happen when the pouch packaged device is stored in rooms with high energy ionizing radiation sources that could generate ozone e.G.Fluoroscopy machines, x ray machines, uv lights, hvac sanitation equipment, etc.The storage conditions for these catheters are specified in the ifu.Corrective actions were implemented to address manufacturing non-conformance.The capa resulted into a voluntary field correction action that instructed customers to return any product which has been stored in the same room as high energy ionizing radiation sources which emit ozone.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|