Model Number FX103T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 05/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: during the investigation, no significant deformation or damage of the valve was determined.Permeability test: a permeability test has indicated that the valve is blocked.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical position.Because the shuntassistant 2.0 is not permeable, a computer control test is not possible.Internal inspection : after dismantling of the valve, deposits were found in shuntassistant 2.0.Results: based on our investigation results, we can determine a blockage in the valve.The determined deposits can be named as the cause for the blockage.Organic matter in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
 
Event or Problem Description
It was reported that a shuntassistant 2.0 (fx103t) was implanted together with a progav 2.0 (#fx410t) during a procedure performed on (b)(6) 2023.According to the complainant, the vale caused an underdrainage.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 1 year.Height: 76 cm.Weight: 6 kg.Gender: female.*same incident/patient as medwatch#3004721439-2024-00160.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNTASSISTANT 2.0 25
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key19587369
Report Number3004721439-2024-00161
Device Sequence Number14329025
Product Code JXG
UDI-Device Identifier04041906501594
UDI-Public4041906501594
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX103T
Device Catalogue NumberFX103T
Device Lot Number20070763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2024
Initial Report FDA Received Date06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age12 MO
Patient SexFemale
Patient Weight6 KG
-
-