This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Unit passed rewind test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.No unexpected motor error alarm noted during testing.Pump history download using thds was successful.However, a47 alarms found on the history file which caused corrupted data.Unable to confirm the reported event date due to the corrupted data.A47 alarms are due to the pump not having power for a long period of time.Pump was cut and open found no moisture/damage on the electronics, battery connector, battery tube, motor, vibrator, keypad flex tail noted during visual inspection.Motor passed motor test.The test reservoir locked in place properly and no anomaly noted.Unit had scratched case, cracked battery tube threads, cracked reservoir tube lip, stained address/serial number label and missing end cap sticker.Unit passed all required test.Motor error alarm was not confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|