• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 030456
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant product: 030449 endo giaii uni ins (lot#: p2f0576s); 030456 endo gia r/or 45 4.8mm x6 (lot#: t1c216x); 030456 endo gia r/or 45 4.8mm x6 (lot#: t1a007x).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a laparoscopic lobectomy procedure, the first firing failed to fire, and three reloads also failed to fire.A new handle and reload were used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the instrument firing knobs were retracted, and the articulation lever was in neutral position.Visual inspection of int ernal components revealed sheared teeth on the firing rack.Functional testing found that the instrument was successfully loaded with a reload.During the firing cycle, a skip was audible in the firing stroke.An access hole was cut into the instrument body for vi sualization of the firing rack.It was reported that the instrument did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur in any of the following circumstances: firing over tissue that is beyond the recommended thickness range.Firing with an obstacle incorporated in the jaws.Attempting to fire a reload that is in interlock.In each circumstance, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19588344
Report Number1219930-2024-02779
Device Sequence Number3354183
Product Code GDW
UDI-Device Identifier10884523002874
UDI-Public10884523002874
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/24/2026
Device Model Number030456
Device Catalogue Number030456
Device Lot NumberT1C216X
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2024
Supplement Date Received by Manufacturer09/09/2024
Initial Report FDA Received Date06/21/2024
Supplement Report FDA Received Date09/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H11
Patient SexPrefer Not To Disclose
-
-