• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 4.5MM(D) 6.5MM(P) 5MM(H); DENTAL HEALING ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 4.5MM(D) 6.5MM(P) 5MM(H); DENTAL HEALING ABUTMENT Back to Search Results
Catalog Number TEEHA465
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Multiple reports are being submitted for this event.
 
Event or Problem Description
It was reported that the abutments placed in site 18 would not seat, kept spinning.
 
Additional Manufacturer Narrative
This report is being submitted to report additional information.Multiple reports are being submitted for this event.The following sections are being reported: d4: expiration date and unique identifier (udi) number h2: if follow-up, what type h3: device evaluated by manufacturer h4: device manufacturer date h6: type of investigation h6: investigation findings h6: investigation conclusions h10: additional narratives/data zimvie received one healing abutment for evaluation.Visual evaluation of the as returned device identified the abutment with signs of use, and slight wear however, the abutment engaged/seats as intended with an in-house zimmer implant during a functional test.No malfunction.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are incorrect techniques used - customer error.Therefore, based on the available information, and functional testing a device malfunction did not occur.The reported event was non-verifiable with all the available information and without return of all devices involved.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional or corrected information to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 4.5MM(D) 6.5MM(P) 5MM(H)
Common Device Name
DENTAL HEALING ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key19588802
Report Number0001038806-2024-01306
Device Sequence Number11598173
Product Code NHA
UDI-Device Identifier00844868045885
UDI-Public(01)00844868045885(17)271003(10)1258558
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K212730
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberTEEHA465
Device Lot Number1258558
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/21/2024
Supplement Report FDA Received Date11/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age39 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-