|
(b)(4).Date sent: 6/22/2024 d4: batch #781c88 b3: only event year known: 2024.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the gcs40b device was received with no apparent damage and with a reload loaded in the device.The reload was received with the drivers partially advanced, washer uncut, with the knife recessed below the reload deck and all staples inside the pockets.The device was tested for functionality with a test reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.In addition, the ledge of the lockout showed slight indentation.The condition of the reload is consistent with the device being partially activated.As a result of the partial firing the lockout was engaged when the device was reopened.Do not fire the instrument unless the closure trigger is properly latched against the handle.The firing trigger must be pulled back completely against the closure trigger to properly fire the instrument.In addition, if the firing sequence is not complete, you could deploy the staples without cutting the washer and forming the staples.It is possible that an attempt was made to close the device with a reload not properly loaded or with a spend reload, this condition would lock the device giving the impression that the device will not close.The echelon contour¿ curved cutter is designed with a feature that prevents closure if a used reload, no reload, or an incorrectly inserted reload is in the instrument.The used or misloaded reload module should then be replaced with a new reload or, if no reload was present, a reload should be loaded in the instrument.After following the above steps, any device which does not close either partially or completely should not be used.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 781c88, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device gcs40b lot number c9d90a and no non-conformances were identified.
|