| Model Number |
CRIT-LINE BLOOD CHAMBER |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
06/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility¿s clinic manager (cm) reported that a blood leak occurred during a patient¿s hemodialysis (hd) treatment due to a crack in the crit line.The patient¿s estimated blood loss (ebl) is unknown.There was no patient injury, adverse event or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.It is unknown if the complaint device is available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.Should additional information become available, the file will be updated accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
A user facility¿s clinic manager (cm) reported that a blood leak occurred during a patient¿s hemodialysis (hd) treatment due to a crack in a crit-line clip (clic) blood chamber.The patient¿s estimated blood loss (ebl) is unknown.There was no patient injury, adverse event or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.It is unknown if the complaint device is available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.Should additional information become available, the file will be updated accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Event or Problem Description
|
|
A user facility¿s clinic manager (cm) reported that a blood leak occurred during a patient¿s hemodialysis (hd) treatment due to a crack in a crit-line clip (clic) blood chamber.The patient¿s estimated blood loss (ebl) is unknown.There was no patient injury, adverse event or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.It is unknown if the complaint device is available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.Should additional information become available, the file will be updated accordingly.
|
| |
|
Search Alerts/Recalls
|