MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. WHITE RELOAD FOR ECHELON 45; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. WHITE RELOAD FOR ECHELON 45; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECR45W
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the unk surgery, after fired, noted the staples malformed.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 6/24/2-24.D4: batch # unk.The device/photo upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 7/17/2024.D4: batch #826a08.D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that one ecr45w reload was received unfired with an open package.The returned reload was loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: tissue thickness should be carefully evaluated before firing any stapler.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.When positioning the stapler on the application site, ensure that no obstructions such as clips, stents, guide wires, etc.Are within the instrument jaws.The event described could not be confirmed as the reload performed without any difficulties noted.No malformed staples were observed.Device history review: a manufacturing record evaluation was performed for the finished device batch number 826a08, and no non-conformances related to the reported complaint condition were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITE RELOAD FOR ECHELON 45
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key19593870
Report Number3005075853-2024-04626
Device Sequence Number11776597
Product Code GDW
UDI-Device Identifier10705036001768
UDI-Public10705036001768
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECR45W
Device Lot Number969A07
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/03/2024
Supplement Date Received by Manufacturer06/25/2024
Initial Report FDA Received Date06/24/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-