| Catalog Number |
151650510 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Adhesion(s) (1695); Muscular Rigidity (1968); Pain (1994); Synovitis (2094)
|
| Date of Event |
06/10/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the chief complaint was painful right tka.The patient was inactive for a long stretch after a severe bout of covid.There is no sign of infection or implant loosening.Surgeon decided to perform synovectomy, removal of scar tissue, and posterior capsular release.There was no sign of infection or implant loosening.Poly was removed and replaced in order to gain access to the posterior knee capsule surgical delay was unknown.Doi; (b)(6) 2019.Dor: (b)(6) 2024.Affected side: right knee.
|
| |
|
Event or Problem Description
|
|
Additional information was received: -the poly was removed in order to provide access to the posterior capsule.You can say there was zero delay or every minute of the case (approx 55 min) was a delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|