|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection.During evaluation, it was confirmed that there was a breakage at the border between the metal of the forceps port connection and the plastic of the actuator.Buckling occurred near the base of the insertion; the stylet could be pulled out, but resistance was felt at the buckling of the insertion during withdrawal.Based on the results of the investigation, no needle breakage was identified, but the needle was crushed, likely due to a bending load on the needle tube.However, a definitive root cause could not be identified.The event can be prevented by following the instructions for use which state: content of the instruction manual (drawing number: rk0385, revision number :03) was confirmed.The instruction manual contains the following descriptions, and it warns against this event.Turn the elevator control lever of the endoscope in the opposite direction of ¿ u to lower the forceps elevator before inserting the instrument.Otherwise, the distal end of the sheath may be blocked by the forceps elevator, which could cause it to extend abruptly from the distal end of the endoscope.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage.It could also damage the endoscope and/or the instrument.Use slow short strokes to insert the instrument into the endoscope channel, do not insert the instrument abruptly.Doing so could cause patient injury, such as perforation, bleeding, or mucous membrane damage.It could also damage the endoscope and/or the instrument.Do not force the instrument if resistance is encountered.Reduce the angulation of the endoscope until the instrument passes smoothly.If excessive resistance is encountered after reducing the angulation of the endoscope entirely, do not use the instrument.Attempting to force the instrument could cause patient injury, such as perforation, bleeding, or mucous membrane damage.It could also damage the endoscope and/or the instrument.Do not coil the insertion portion with a diameter of less than 20 cm.This could damage the insertion portion.If the insertion portion has any malformations like kinks, bends, or cracks, do not use the instrument.Otherwise, unexpected perforation, bleeding, or membrane damage may occur.Be sure to check that the forceps elevator is down before withdrawing the instrument from the endoscope.When the needle slider is withdrawn with the forceps elevator up, the needles distal end cannot be completely retracted into the sheath.When the incompletely retracted needles distal end is withdrawn from the endoscope, the needles distal end extending from the sheath may cause damage to the endoscopes instrument channel.The handle section and sheath should be kept straight, when withdrawn from the endoscope.Otherwise, the sheath and/or needle tube may be bent, negatively affecting specimen extraction.Olympus will continue to monitor field performance for this device.
|