| Model Number |
2ACH20 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226)
|
| Date of Event |
06/20/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a cryo ablation procedure, the patient became hypotensive and the cardiac shadow disappeared on imaging.A pericardial tamponade was then confirmed.The case was aborted.The patient was not under general anesthesia.The pericardial tamponade was treated.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id afapro28 ; product type: balloon catheter; product id 4fc12 ; product type: sheath.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was later reported that the patient was rushed to the surgical department for emergency treatment but the specific intervention is unknown.The patient's hospitalization was extended and is still currently hospitalized.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the 2ach20 mapping catheter with lot number 8865194 and data files were returned and analyzed.The patient file was received and recorded on the reported date of the event.The patient file showed five applications were performed using a catheter identified as afapro28 of lot number 22002.The patient file didn't show any system notice on the reported date of the event.Visual inspection of the loop segment area showed the loop was intact with no apparent issues.No damage was observed along with the tip/loop section of the achieve mapping catheter.Visual inspection of the pebax tubing area showed the pebax tubing was intact with no apparent issues.No damage was observed along with the pebaxtubing section of the mapping catheter.Visual inspection of the electrode(s) showed the electrode(s) was intact with no apparent issues.All electrodes exist on the loop section and no cosmetic issue or anomalies were identified.Visual inspection of shaft segment area showed the shaft was intact with no apparent issues.No kin k or any other damage was observed along with the shaft of the mapping catheter.Visual inspection of the introducer showed the introducer was intact with no apparent issues.No damage or any other issue was observed along with the introducer.Visual inspection of the lemo connector showed the lemo connector was intact with no apparent issues.No damage or any other issue was observed along with the lemo connector.The functional test was performed using a multimeter.The mapping catheter was connected to the test cable.The continuity and impedance measurement between the electrodes and the other side of the cable showed the electrode's continuity and impedance to the cable are normal.In conclusion, the reported cardiac tamponade and hypotension occurred during the procedure and could not be confirmed through data or product analysis.There is no indication of a relation of the adverse event to the performance and malfunction of the product.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|