MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3542757
Medical Device Problem Codes Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); Granuloma (1876)
Date of Event 01/01/2016
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 61-year-old caucasian female patient underwent a breast augmentation revision with two 275cc mentor memorygel breast implants and experienced bilateral capsular contracture (baker grade 4) post-operatively, which was diagnosed with an office visit.As a result, the patient is to undergo bilateral explantation on (b)(6) 2024.This report is for the left side device.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.Section d6b.Explantation date: (b)(6) 2024.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On october 15, 2024, mentor received additional information reporting that the patient also experienced a rupture, which was diagnosed via mammogram.Reason for device explant and/or reoperation: capsular contracture and rupture.D4: udi: udi unavailable, product made prior to udi compliance date.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Correction: section b1 for the reported product problem (rupture) should have been selected in the previous report.The box has been checked on this report.On december 05, 2024, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.On december 06, 2024, mentor received additional information regarding the patient's event.It was reported that the patient also experienced breast pain and granuloma on the left side.It was reported in the operative report that multiple extracapsular silicone granulomas" throughout the left breast.The patient's replacement device was reported to be a 225cc mentor memorygel breast implant (catalog number 3502251bc) with serial number (b)(6).Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On january 07, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned sample revealed that the gel siltex mod-rnd 275cc breast implant had two areas of siltex cracking on the posterior view.In addition, two tears (a and b) were noted on the posterior view within the areas of siltex cracking each tear, and the measurement of the tears were 1.4 cm and 2.1 cm respectively.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.Granulomas are noncancerous nodules that are a common tissue reaction to the presence of a variety of foreign materials, such as silicone.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key19602508
Report Number1645337-2024-07565
Device Sequence Number3433755
Product Code FTR
UDI-Device Identifier00081317000662
UDI-Public00081317000662
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year1997
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2002
Device Catalogue Number3542757
Device Lot Number158907
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/05/2024
Supplement Date Received by Manufacturer10/15/2024
12/05/2024
01/07/2025
Initial Report FDA Received Date06/24/2024
Supplement Report FDA Received Date11/04/2024
12/26/2024
01/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/1997
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-