• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER Back to Search Results
Catalog Number 998504
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that when opened the product the customer found that the manometer on the eagle inflation device was damaged in the packaging.
 
Additional Manufacturer Narrative
The reported event is confirmed cause unknown.Visual evaluation noted one inflation device was returned to atrion inside its tray.Upon receipt of the complaint device a visual inspection was conducted.During the visual inspection, it was noted that the gauge was broken from the inflation device housings at the glue plug area.Additionally, a crystalized substance was found on various parts of the inflator body and the gauge when the device was received.The glue plug threaded section of the device was broken off from the device and remained attached to the gauge threads.The device exhibits the atrion cts mark, indicating 100 percentage functional verification at the time of manufacture and the gauge needle is within the zero position.The cts testing includes pressure leakage, pressure decay, vacuum capability, gauge accuracy.Functional evaluation noted functional testing could not be performed on the inflation device due to the gauge being detached from the inflation device body.Root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be inappropriate package design.However, there was insufficient information to confirm this potential root cause.A dhr review did not show any problems or conditions that would have contributed to the reported event.Labelling review is not required as labelling would not have prevented the reported event.Udi format updated with available product information per gudid.
 
Event or Problem Description
It was reported that when opened the product the customer found that the manometer on the eagle inflation device was damaged in the packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® BALLOON DILATION CATHETER
Common Device Name
BALLOON DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19604478
Report Number1018233-2024-03728
Device Sequence Number7326230
Product Code EZN
UDI-Device Identifier10801741052672
UDI-Public(01)10801741052672(11)220419(17)250228(10)BMGQAT03
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K050875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue Number998504
Device Lot NumberBMGQAT03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2024
Supplement Date Received by Manufacturer09/19/2024
Initial Report FDA Received Date06/25/2024
Supplement Report FDA Received Date09/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-