| Model Number |
7825 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer from united kingdom (uk) reported that the contour next one meter purchased in uk from amazon website was displaying the units of measurement accompanied with the blood glucose readings in mg/dl instead of mmol/l.There was no allegation of an adverse event.The device is not expected to be returned for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in sections a2 and a4 as the customer's age and weight were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history and distribution records were reviewed for the suspected contour next one meter with serial #(b)(6).It was found that the meter was programmed to display the units of measurement in mg/dl and was intended for and distributed in the us market.The meter was not configured to be distributed in the uk market.The issue occurred outside of ascensia's control.
|
| |
|