• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE ALCON CLOSURE SYSTEM (ACS), NYLON; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE ALCON CLOSURE SYSTEM (ACS), NYLON; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE Back to Search Results
Catalog Number 8065710401
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that an ophthalmic surgical suture was broken during trabeculectomy surgery.The surgery was completed on the same day after replacing with another new suture.It was also reported that the issue had happened four to five times in the same surgery and time consumed to open the new sutures.No patient impact was reported.
 
Additional Manufacturer Narrative
Additional information was provided in sections d.9., h.3., h.6., and h.11.13 unopened sutures, in a pouch, in a blister, in a box were received for the report of suture breaks.Per sampling plan, 10 samples were visually inspected and found to be conforming.A dimensional inspection was done for suture diameter and all 10 samples were found to be conforming.Functional testing for suture strength was performed and all 10 samples were found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned unopened samples were found to be visually, dimensionally and functionally conforming, therefore the suture breaks as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.However since the actual complaint samples were not returned, a root cause cannot be determined for the complaint as described by the customer.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.Suture material samples are tested at incoming for tensile strength and diameter and visually inspected for any defects such as discoloration and frays.Sutures are also 100% inspected by trained operators for discoloration and frays and for proper attachment during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALCON CLOSURE SYSTEM (ACS), NYLON
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key19605534
Report Number2523835-2024-01540
Device Sequence Number6434017
Product Code GAR
UDI-Device Identifier00380657104017
UDI-Public00380657104017
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K930517
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065710401
Device Lot Number153MF1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2024
Supplement Date Received by Manufacturer09/13/2024
Initial Report FDA Received Date06/25/2024
Supplement Report FDA Received Date10/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-