|
Additional information (05-aug-2025): siemens healthineers has confirmed, through an internal investigation of customer complaints, the potential for falsely depressed patient, quality control (qc), and/or calibration results.The observations have been isolated to the initial replicate(s) of a freshly punctured atellica ch ucfp reagent well that has been stored onboard the system.The issue is intermittent and will not occur in all packs or wells.The issue may be observed with all sample types (urine and cerebrospinal fluid) for any atellica ch ucfp reagent lot on an atellica ch or ci analyzer.The maximum negative bias observed during the investigation was -19.0 mg/dl (-190 mg/l).Us customers were notified through urgent medical device correction (umdc letter achc25-07.A.Us), titled ¿potential for falsely depressed results with the atellica ch ucfp assay¿ and ous customers were identified through urgent field safety notice (ufsn letter, achc 25-07.A.Ous) of the potential for falsely depressed patient, quality control (qc), and/or calibration results.The umdc and ufsn provide instructions for customers to perform qc on each well.The customer is operational.No further evaluation is required.In section h6, the investigation finding and investigation conclusion codes were updated.Initial mdr 2432235-2024-00141 was filed on 25-jun-2024.
|