(b)(4).Date sent: 6/25/2024.D4: batch # unk.Investigation summary : the product was returned to ethicon endo-surgery for evaluation.Visual inspection was conducted on the returned reload.Visual analysis of the returned sample revealed that the ecr45w reload was returned unfired with the reload pan partially dislodged from the right proximal side.In addition, 3 double drivers and 6 single drivers missing, making the reload non-functional.No functional test was performed due to the condition of the reload.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached on what may have caused the reload pan to dislodge.A manufacturing record evaluation was performed for the finished device batch number t5ak02, and no non-conformances were identified.
|