| Catalog Number |
07.01702.023 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during final tightening of a c4/c7 fixation, an abnormal sound was observed, and found that two virage screws' housings were fractured at the welds.There was a 30-minute delay to remove the broken screws and replace them with new screws.There was no impact on the patient.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3012447612-2024-00136.
|
| |
|
Event or Problem Description
|
|
It was reported that during final tightening of a c4/c7 fixation, an abnormal sound was observed, and found that two virage screws' housings were fractured at the welds.There was a 30-minute delay to remove the broken screws and replace them with new screws.There was no impact on the patient.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d9, g1, and h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the weld has fractured on the lower housing of the screw, resulting in the screw disassembling.A dimensional inspection of the weld depth was performed on a smartscope (id 0444 cal.Due 5/16/2025).The measurements are depicted in the attached photos and were set at the minimum allowable depth of.010".The weld depth exceeds the minimum requirement per drawing 07.01702.000 rev.D.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to excessive torque applied to the closure top during final tightening.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|