|
(b)(4).The actual device was not returned; however, the customer provided one photo for analysis.The complaint of a catheter kinked was able to be confirmed by the photo.The photo shows a used arterial catheter with a small kink near the hub, however, a complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "warning: care should be exercised that the catheter is not inadvertently kinked at the hub area when securing catheter to the patient as this may result in catheter damage, breakage and loss of arterial monitoring capabilities." the customer report of a kinked catheter was confirmed by visual inspection of the customer supplied photo.The photo shows a used arterial catheter with a small kink adjacent to the hub, however, full complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|