Catalog Number 59-BP-6545
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/30/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D4 lot number - unknown h4 device manufacture date - unknown without lot identification h3 device evaluation - the complaint device was ultimately implanted and cannot be returned for evaluation.Without the opportunity to examine the complaint product no conclusions can be drawn regarding the root cause of the reported malfunction.Review of device history records and lot specific complaint history review were not possible without lot identification.Two-year complaint history review did not reveal a trend for reports of this nature for this part number.This report is number 2 of 2 mdrs filed for the same event (reference (b)(4).
 
Event or Problem Description
It was reported that a procedure was performed in australia on (b)(6) 2024, utilizing the reform mc midline cortical screw system.While inserting a reform ti modular polyaxial tulip assembly (39-mt-0401) assembled on a ø6.5 x 45mm midline cortical bone screw (59-bp-6545) the tulip came off of the screw when the driver was removed.The tulip dropped into the wound and was retrieved using forceps.The tulip was then snapped back onto the screw shank and the procedure was completed with only a short delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFORM MIDLINE CORTICAL SCREW SYSTEM
Common Device Name
PEDICLE SCREW
Manufacturer (Section D)
PRECISION SPINE, INC
2050 executive drive
pearl MS 39208
Manufacturer (Section G)
PRECISION SPINE, INC
2050 executive drive
pearl MS 39208
Manufacturer Contact
mike dawson
2050 executive drive
pearl, MS 39208
6014204244
MDR Report Key19608477
Report Number3005739886-2024-00030
Device Sequence Number20884107
Product Code NKB
UDI-Device Identifier00815362026342
UDI-Public00815362026342
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K173130
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number59-BP-6545
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/30/2024
Initial Report FDA Received Date06/25/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-