H6: corrected the following: health effect - impact code, component code, type of investigation, investigation findings, investigation conclusions.Manufacturer narrative updated: based on the available information, the patient involved in the incident meets the following risk criteria for early prosthesis wear and/or osteolysis: combination of largest available femoral head and/or thinnest available acetabular liner was used, and implanted with a component having a shelf age of greater than 2 years.The most likely cause for the reported revision due to prosthesis wear is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.
|