Product complaint (b)(4).Investigation summary: it was reported, that the trials all have scratches and nicks out the plastic.Hospital will not allow them to be used in cases.The event did not happen in surgery.The product was returned to depuy synthes for evaluation.Visual inspection revealed, scratches from the saw blade excursion on one flange of the attune cr fem trial sz 4 rt.Additionally, the color-coded indicator of the trial was severely faded.The scratches observed, are consistent with the device coming in contact with the saw blade while trailing.Care should be taken to avoid saw blade excursion into the femoral trials or implants.The fading condition was identified, as an end of life indicator.Damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use.The device must be properly inspected, prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions and inspection procedures.A dimensional inspection was not performed, since it was not applicable to the complaint condition.A functional test was not performed, since it was not applicable to the complaint condition.The overall complaint was confirmed.As the observed, condition of the attune cr fem trial sz 4 rt would contribute to the complained device issue.Based on the investigation findings, it has been determined, that no corrective and/or preventative action is proposed.There is no indication, that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H6: component code: appropriate term/code not available.(g07002) used to capture no findings available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
|