One opened trocar assembly in a small parts tray (smpt) was received for the report.Sample was visually inspected and found to be nonconforming, damaged blade tip/bent tip and point broken off tip.Surgical material was observed on blade.Penetration testing could not be performed due to the damage of the sample.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.The exact root cause could not be determined from the investigation performed.The damage to the returned sample is consistent with damage that can occur when the blade contacts a hard surface such as the protective cap, improper handling, or contact with another instrument during surgery or set-up.The damage seen on the returned complaint samples could have contributed to the customers reported issue of cutting of the trocar was poor during surgery.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All trocar blades are 100% inspected.Any nonconformances, such as damaged tips, are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|