| Model Number |
G0404-32P01C-1 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Unintended Radiation Exposure (4565)
|
| Date of Event |
05/31/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
According to the reporter, during an orthopedic right total knee revision, the surgeon packed the bone lumen with the raytec device.During the case, they were missing a sponge, so they scanned the patient and the scan was clear.Therefore, they assumed that the patient was clear and it was possibly in the trash.At the end of the case, they performed an x-ray and found the sponge in the bone lumen.The next day they brought the patient back to the operating room to finish the procedure.After retrieving the sponge, the pouch was damaged, and they were able to pull the chip out that also looked to be damaged, the chip fell out and was not resonating when scanned outside the body.Photo was submitted showing the chip of the device was broken/detached from the sponge and sponge fiber coming off.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: 01-0046 blair port wand x (lot #: pt00038502).01-0043 console, model 200x (serial #: (b)(6).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the gauze was used and was ripped.The chip was out of its housing.The part of the chip was melted.It was reported that the component disengaged and was linting.The reported issue were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|