| Model Number |
0275 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
Bradycardia (1751)
|
| Date of Event |
06/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead presented noise at the beginning of a mediated tachycardia (pmt) episode which cause pacing inhibition.Boston scientific technical services (ts) advised pmt appeared to be sinus tach at maximum tracking rate (mtr).Isometrics and pocket manipulation testing was performed which did not reproduce the noise on the rv rate/sense.The patient took deep breaths, and some noise was visible on the rate/sense.Ts noted the noise appeared to be diaphragmatic noise and not lead integrity.Rv measurements remain stable and within normal limits.Ts confirmed the limited presence of oversensed noise but not for long periods and recommended increasing the rv sensing measurements or leave it as it is.This lead remains in service no adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|