Model Number 0275
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Code Bradycardia (1751)
Date of Event 06/11/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this right ventricular (rv) lead presented noise at the beginning of a mediated tachycardia (pmt) episode which cause pacing inhibition.Boston scientific technical services (ts) advised pmt appeared to be sinus tach at maximum tracking rate (mtr).Isometrics and pocket manipulation testing was performed which did not reproduce the noise on the rv rate/sense.The patient took deep breaths, and some noise was visible on the rate/sense.Ts noted the noise appeared to be diaphragmatic noise and not lead integrity.Rv measurements remain stable and within normal limits.Ts confirmed the limited presence of oversensed noise but not for long periods and recommended increasing the rv sensing measurements or leave it as it is.This lead remains in service no adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19615914
Report Number2124215-2024-39285
Device Sequence Number20409632
Product Code NVY
UDI-Device Identifier00802526531224
UDI-Public00802526531224
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P910073/S077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/15/2025
Device Model Number0275
Device Catalogue Number0275
Device Lot Number422357
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2024
Initial Report FDA Received Date06/26/2024
Date Device Manufactured02/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
-
-