Product analysis #(b)(4):equipment used: video inspection system (m-85519), ruler (m-83361), camera (panasonic lumix dmc-zs5 as found condition: the axium prime device was returned for analysis within a shipping box; within a sealed plastic biohazard pouch; within a dispenser coil and within an introducer sheath.The unknown micro catheter used in the event was not returned for analysis.Two other axium prime devices were also returned and will be addressed in pli-10 and pli-20.Visual inspection/damage location details: no damages or irregularities were found with the introducer sheath.No damages or irregularities were found with the axium prime pusher.The axium prime implant coil was found stretched and damaged within the introducer sheath with the polypropylene filament broken.Testing/analysis: n/a conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil stretch¿ was confirmed.Possible causes of failure are lack of hydration, user does not maintain continuous flush, tortuous anatomy, coil not retracted in a one-to-one motion with the implant during repositioning, damaged micro catheter, incompatible micro catheter, pushwire rotation or user advances the coil against resistance.Customer reported devices were prepared per ifu and vessel tortuosity was normal.As no other information was provided, likely cause could not be determined.As the unknown micro catheter used in the event was not returned for analysis, any contribution of the micro catheter towards the coil stretch could not be determined.As the model and lot numbers were not reported, compatibility could not be assessed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|