|
|
| Catalog Number |
VR214 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained: please clarify when the event occurred: when the returned product was observed, it was found that degradation has begun.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.H6 component code: g07002 ¿ device not returned.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent an unknown procedure on an unknown date, and suture was used.The suture was broken at the root of the needle.It was not a control release needle.There were no adverse consequences to the patient.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 8/16/2024.H6 component code: g07002 no device problem found.Corrected information: h6 investigational codes, h3 investigation summary - the product was returned to ethicon for evaluation.Two unopened samples that pertain to product code vr214 were received for analysis.Additionally, a photo was provided, with the same condition as the received sample.During visual inspection of the samples, the packets were opened.It was observed that the winding former contained a suture.However, the suture cannot be dispensed due to it has begun with the degradation process; per the sample condition, the functional test cannot be performed.The foil was visually inspected, and excessive wrinkles and holes in the cavity were observed.The reported event is confirmed.Due to the damages found on the packaging, a possible cause for this condition is due to improper handling during transit or storage.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|