• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC OPTRELL MAPPING CATHETER WITH TRUEREF TECHNOLOGY; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC OPTRELL MAPPING CATHETER WITH TRUEREF TECHNOLOGY; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Catalog Number UNK_OPTRELL MAPPING CATHETER
Medical Device Problem Codes Entrapment of Device (1212); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Mitral Valve Insufficiency/ Regurgitation (4451)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This complaint is from a literature source.The following literature cite has been reviewed: usman bhatti, ethan hacker, jonathan m.Burnett, sunny s.Po.(may 2024).Deer stuck in a fence; entrapment of high-density grid mapping catheter (optrell) in mitral subvavlvular apparatus.Heart rhythm society.Po-02-027.Https://doi.Org/10.1016/j.Hrthm.2024.03.907.No device was received for analysis at the time of submission of the initial 3500a.Since the product was not returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors could be made.Device history record (dhr) review cannot be conducted because no lot number was provided by the customer.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref #: (b)(4).
 
Event or Problem Description
This complaint is from a literature source.The following literature cite has been reviewed: usman bhatti, ethan hacker, jonathan m.Burnett, sunny s.Po.(may 2024).Deer stuck in a fence; entrapment of high-density grid mapping catheter (optrell) in mitral subvavlvular apparatus.Heart rhythm society.Po-02-027.Https://doi.Org/10.1016/j.Hrthm.2024.03.907.Objective/methods/study data: background: a 55 year old male with ischemic cardiomyopathy was referred to electrophysiology lab for vt ablation.A novel grid catheter, optrell (biosense webster, ca) was used to map lv via retro-aortic approach.Objective(s): first known case of entrapment of the optrell catheter in the mitral apparatus.Additionally we describe our multi-modality approach to rescue the situation and avoid emergent surgery.Method(s): while mapping the lv, the catheter became difficult to manipulate, we noticed narrowing of the splines around the middle like in a noose.Ice and fluoroscopy verified this concern.Gentle rotation, advancement and retraction failed to free the catheter.Attempt to use ablation catheter via transseptal to push the optrell apically failed.Tee showed catheter entangled in the chordae leading to severe mitral regurgitation.We then utilized the retrospective review feature (cartoreplay) of mapping system (carto 3; biosense webster, ca) and were able to visualize the exact catheter movements leading to entrapment.We saw that catheter entered lv directing to basal lateral wall.Attempting to map the basal anteroseptal wall, a clockwise rotation & then complete retro-flexion of optrell was performed, resulting in the splines getting pinched in middle.Result(s): fortunately, we were successful in performing all these maneuvers in the exact stepwise but opposite fashion to untangled the catheter from the subvalvular apparatus and retract it back to the aorta.Tee afterwards showed no valvular complication and resolution of mitral regurgitation.Conclusion(s): use of the grid mapping catheter may present a potential hazard in patients with lax subvalvular apparatus.Although the bi-directional deflection of optrell catheters increases maneuverability, caution should be taken if catheter manipulation becomes difficult, especially in a retro-aortic approach.Lot, model and catalog number are not available, but the suspected biosense device possibly associated with reported adverse events: optrell.Concomitant biosense webster devices that were used in this study: carto concomitant non-biosense webster devices that were also used in this study: n/a.Adverse event(s) and provided interventions possibly associated with unidentified optrell: qty 1 ¿ 55 year old male with ischemic cardiomyopathy experienced severe mitral regurgitation (mitral valve incompetence) (serious injury) as a result of (medical device entrapment ¿ excessive manipulation required).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTRELL MAPPING CATHETER WITH TRUEREF TECHNOLOGY
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key19618157
Report Number2029046-2024-02098
Device Sequence Number6963593
Product Code DRF
UDI-Device Identifier10846835023015
UDI-Public10846835023015
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberUNK_OPTRELL MAPPING CATHETER
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/28/2024
Initial Report FDA Received Date06/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK_CARTO 3.
Outcome Attributed to Adverse Event Other;
Patient Age55 YR
Patient SexMale
-
-