BOSTON SCIENTIFIC CORPORATION CHOICE PT; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 2040
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the balloon was difficult to remove and almost detached from the shaft and the guidewires were partially detached.The target lesion was located in the left anterior descending artery.The balloon was inflated for multiple inflations at 14 to 16 atm for 12 to 14 seconds.A 15mmx3.00mm wolverine coronary cutting balloon was selected for use.Upon removal, the wolverine balloon would not come back through the choice pt floppy guidewire upon deflation.The physician tried to pull double negative and took an empty 20cc syringe and pulled negative.Neither of these techniques worked.The physician then inflated the wolverine to 2 atm and deflated but it still would not come through to the guidewire.The physician then placed another choice pt floppy guidewire next to the wolverine with a 2.5mmx 12mm emerge balloon catheter onto the wire.Physician inflated next to the wolverine and was finally able to nudge the balloon enough to fully retract it out of the body.Physician said it was still very difficult to retract.Upon investigation, it was noted that the two wires were both frayed and partially detached and the wolverine seemed to almost be detached from the shaft.The device was removed as one unit together with the guidewire, and procedure was completed using another device.No further complications were reported, and patient is stable post procedure.
 
Event or Problem Description
It was reported that the balloon was difficult to remove and almost detached from the shaft and the guidewires were partially detached.The target lesion was located in the left anterior descending artery.The balloon was inflated for multiple inflations at 14 to 16 atm for 12 to 14 seconds.A 15mmx3.00mm wolverine coronary cutting balloon was selected for use.Upon removal, the wolverine balloon would not come back through the choice pt floppy guidewire upon deflation.The physician tried to pull double negative and took an empty 20cc syringe and pulled negative.Neither of these techniques worked.The physician then inflated the wolverine to 2 atm and deflated but it still would not come through to the guidewire.The physician then placed another choice pt floppy guidewire next to the wolverine with a 2.5mmx 12mm emerge balloon catheter onto the wire.Physician inflated next to the wolverine and was finally able to nudge the balloon enough to fully retract it out of the body.Physician said it was still very difficult to retract.Upon investigation, it was noted that the two wires were both frayed and partially detached and the wolverine seemed to almost be detached from the shaft.The device was removed as one unit together with the guidewire, and procedure was completed using another device.No further complications were reported, and patient is stable post procedure.
 
Additional Manufacturer Narrative
B5 - describe event or problem: updated.The device was not returned for evaluation.However, an analysis was concluded based on the media attached to the parent record showing a detached guidewire.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHOICE PT
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19619115
Report Number2124215-2024-38978
Device Sequence Number6197670
Product Code DQX
UDI-Device Identifier08714729177005
UDI-Public08714729177005
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K143587
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2040
Device Catalogue Number2040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2024
Supplement Date Received by Manufacturer07/05/2024
Initial Report FDA Received Date06/26/2024
Supplement Report FDA Received Date07/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-