| Model Number |
457 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint was able to be confirmed based on information received from the end user.Root cause is determined to be accidental user error in regards to power settings/activation timing.The device was being used at 40w and being held active for a few minutes at a time.Due to the extended activation timing at a higher power setting, the wire broke due to thermal/mechanical stress.If additional information is obtained that is pertinent to the investigation or provides additional details a follow-up report will be provided.Until such time, this can be seen as the final report.
|
| |
|
Event or Problem Description
|
|
The complainant alleges, "the 457 loop electrode with lot 0223z broke.".
|
| |
|