Model Number GCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Discomfort (2330)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section a3 and a6: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section b3 - date of event: unknown, as information was requested but not provided.Section d6a - implant date: unknown, as information was requested but not provided.Section e1 - telephone number: (b)(6).Device evaluation: product testing could not be performed since the device was not returned for evaluation.Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no additional complaints were received for this po.No escalation required.Conclusion: a product deficiency has not been identified; therefore, no additional corrective actions have been initiated.Attempts have been made to obtain the missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the patient is unsatisfied with the results following intraocular lens (iol) implantation.Through follow-up, we learned that the patient's left eye uncorrected visual acuity was +0.1 before the surgery and +0.5 after the surgery.The iol was explanted and replaced with a johnson and johnson ar40e model iol.The patient did not present any other visual problems or injuries.No other medical interventions were performed, and no issues with the iol were reported.This event was not debilitating to the patient.No further details were provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTLOAD
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key19620129
Report Number3012236936-2024-000167
Device Sequence Number12332839
Product Code HQL
UDI-Device Identifier05050474620551
UDI-Public(01)05050474620551(17)260329
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/29/2026
Device Model NumberGCB00
Device Catalogue NumberGCB0000175
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2024
Initial Report FDA Received Date06/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexFemale
Patient Weight100 KG
Patient EthnicityNon Hispanic
-
-