• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number ATG80126
Medical Device Problem Code Missing Information (4053)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/28/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to an angioplasty procedure, it was noticed that the three codes with the information allegedly does not have barcodes.There was no patient contact.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 10/2026).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.One electronic photo was provided for review.The photo shows a portion of an atlas gold label with lot and product information matching the reported information in trackwise.No barcode can be seen on the label.The atlas gold labeling label artwork document was reviewed, and the updated label including a barcode on the device was implemented on 10/31/2023 at the manufacturing facility.The previous label (baw1402400 rev.2) prior to this date did not include a barcode.Therefore, the investigation is unconfirmed for the reported missing barcode as the product labeling contains all correct information based on the date of manufacture.The barcode was added to the labeling after the manufacture date of this lot.The cause of the reported event is likely this change in labeling being unknown to the reporter.Labeling review: atlas gold label artwork document was reviewed.Revision 3 was the first revision to include a barcode on the outer product labeling.The second revision did not contain this barcode on the label.The third revision was implemented at the manufacturing facility on 10/31/2023.Per the device device history records review, this lot was manufactured and released on 10/30/2023.Therefore, this lot would have used revision 2, which did not contain a barcode.As a result of the labeling review, it can be determined that no labeling issue can be confirmed as the product labeling contains all correct information based on the date of manufacture.H10: d4 (expiry date: 10/2026), g3, h6 (method).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that prior to an angioplasty procedure, it was noticed that the three codes with the information allegedly does not have barcodes.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19621511
Report Number2020394-2024-00928
Device Sequence Number18264288
Product Code DQY
UDI-Device Identifier00801741060717
UDI-Public(01)00801741060717
Combination Product (Y/N)N
Initial Reporter CountryIR
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberATG80126
Device Lot Number93XH0141
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/28/2024
Supplement Date Received by Manufacturer08/06/2024
Initial Report FDA Received Date06/27/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-