| Model Number |
MENISCAL CINCH |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 7/19/2023, it was reported by an arthrex employee via email that an ar-4500 meniscal cinch did not deploy the second anchor.The surgeon attempted to deploy the second anchor three times but could not.The procedure was completed using a product from another manufacturer.This was discovered during an aclr and mencius repair procedure on (b)(6) 2023.Additional information received on 7/27/2023: the first anchor was removed; however, the case was delayed nearly thirty minutes, and additional anesthesia was administered to the patient.Nothing broke off inside the patient.The same issue occurred with two other ar-4500 meniscal cinch.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is not confirmed.Visual inspection confirmed that the implants were deployed and there was no presence of sutures.Functional testing could not be conducted because the device was missing trocar #1, #2, and the depth stop tab.No problem found.
|
| |
|