| Model Number |
PCF-HQ190I |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The legal manufacturer's investigation is ongoing, this report will be supplemented when new and relevant information becomes available.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the flex video scope exhibited the air/water, air water tube and jet tube had foreign material.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information added on fields: d8, h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.The legal manufacturer reviewed the customer provided the cleaning disinfection and sterilization (cds) processes where no obvious deviations from instructions for use (ifu) were identified.Based on the results of the investigation and the information provided, it could not be definitively determined what the foreign material was in the subject device.There was no damage to the area where the foreign material was detected, and reprocessing was performed according to the instructions for use (ifu).Therefore, the cause of the material remaining in the device could not be determined.The event can be detected/prevented by following the instructions for use (ifu): ifu states that detection method in gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection.Ifu states that preventive measure in gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope.Three attempts were made to obtain additional information regarding the reported event, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|