| Model Number |
FIBERLOOP, #2, BLUE |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament reconstruction with semitedinosus tendon the device fell apart.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint allegation is not confirmed as the complaint device was not received for evaluation.Four sealed packaged ar-7234 serial/batch number 15068356 were received for evaluation.The complaint allegation device was not received for evaluation.Two packages were open, and upon visual inspection, the suture needle and assembly were free of defects.According to drawing c1283, the nitinol loop must withstand a 5 lb straight pull.When the device¿s documentation was reviewed, no issues were found.The device passed the pull functional test.Functional testing was performed by applying manual tension to the suture to replicate suture tensioning, and no issues were noted.The needle was not observed to loosen or detach.No problem was found with the sealed package received devices.The most likely cause of the reported failure is a user error.Direction for use.Dfu-0222-eor2_fmt_en.G.Precautions.2.In handling this or any other suture material, care should be taken to avoid damage from handling.Avoid crushing or crimping damage due to application of surgical instruments such as forceps or needle holders.
|
| |
|
Search Alerts/Recalls
|
|
|