| Catalog Number |
706.52 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that end of an ophthalmic surgical scissors snapped off inside the right eye of the patient and was retrieved during cutter lensectomy, pars plana vitrectomy (ppv), removal of silicon oil (roso), retinectomy, endolaser, oil, membrane dissection surgery.It was also reported that enlargement of scleral wound was required in order to remove broken scissor fragment.The surgery was completed on the same day after replacing with another scissors with same gauge.No patient impact was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in sections h.6., and h.11.A review of the device history record traceable to the reported lot number indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A review of the database for the non-conformance events and the complaints showed that the given lot number was not concerned by any deviation and that there are no additional complaints associated it.The concerned sample was not received at the investigation site for evaluation.Therefore, no further assessment is possible and the investigation is concluded without identifying the root cause.A sample was not received at the investigation site.Therefore, the root cause for the customer complaint issue cannot be determined conclusively.The exact root cause for the customer's reported event is unknown, therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.If new information is received, this case will be re-opened and the input will be considered and reflected.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|