DEPUY ORTHOPAEDICS INC US 11/13 S-ROM 40MM M SPEC+3; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 136543500
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Foreign Body Reaction (1868)
Date of Event 04/16/2014
Type of Reportable Event Serious Injury
Event or Problem Description
Patient had a failed acetabular component on metal-on-metal total hip with evidence of trunnion corrosion and no evidence of infection.During the operation, there was no evidence of pus and metallosis but there was some mild staining metallic looking in the inferior aspect of the acetabulum.There was some mild staining behind the liner but very minimal.The stem was grossly loose, had rotated into retroversion and was grown over with bone.The head had some corrosion of the taper.The sleeve was then cut using sharp dissection because it was scarred down.Doi: (b)(6) 2009.Dor: (b)(6) 2014 left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The adverse symptoms alleged and product code reported is associated with the depuy metal on metal articulation.A complaint database search and/or device manufacturing (dhr) reviews will not be performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11/13 S-ROM 40MM M SPEC+3
Common Device Name
S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19632343
Report Number1818910-2024-14068
Device Sequence Number20692261
Product Code LPH
UDI-Device Identifier10603295033776
UDI-Public(01)10603295033776(10)2832562(17)140131
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2014
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2014
Device Catalogue Number136543500
Device Lot Number2832562
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/18/2024
Supplement Date Received by Manufacturer09/11/2024
Initial Report FDA Received Date06/28/2024
Supplement Report FDA Received Date09/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/04/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINNACLE MTL INS NEUT40IDX58OD.; PINNACLE SECTOR II CUP 58MM.; S-ROM*SLEEVE PRX ZTT, 18D-LRG.; SROM STM STD 36+12L 13X18.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
-
-