• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KLS MARTIN SE & CO. KG LEVEL ONE CMF; SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KLS MARTIN SE & CO. KG LEVEL ONE CMF; SCREW Back to Search Results
Model Number 25-882-07-91
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that screws had become loosened from patient's compromised bone.They were removed.
 
Additional Manufacturer Narrative
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at kls se.During the investigation the product lot number was not identified; therefore, the device history records were not reviewed.Based on the information provided the results conclude that the most likely root cause is patient related.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
SCREW
Manufacturer (Section D)
KLS MARTIN SE & CO. KG
kolbinger strasse 10
muehlheim/donau, de 78570
GM  78570
Manufacturer (Section G)
KLS MARTIN SE & CO. KG
kolbinger strasse 10
muehlheim/donau FL 78570
Manufacturer Contact
melissa bachorski
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key19632351
Report Number9610905-2024-00032
Device Sequence Number7437189
Product Code MQN
UDI-Device Identifier00888118047854
UDI-Public(01)00888118047854
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K032442
Exemption Number2017029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25-882-07-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2024
Initial Report FDA Received Date06/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexMale
-
-