The complaint allegation is confirmed.Upon visual inspection, no broken parts were observed on the device.However, the power pick's inner shaft does not retract when the lever is actuated.Functional testing with the console and the handpiece found that the inner shaft doesn't move because of the break.The most likely cause(s) of the reported failure is user error, such as prying/leveraging the device.Dfu-0215-eor0_fmt_en.F.Precautions.6.Do not allow the rotating portion of any device to touch any metallic object, such as a cannula or arthroscope, during use.Damage to both devices is likely.Damage to the device can range from a slight distortion or dulling of the blade/burr edge to actual fracture of the tip in vivo.If such contact does occur, inspect the tip.If there are cracks, fractures or dulling, or if there is any other reason to suspect a blade is damaged, replace it immediately.7.Excessive "side-loading" on the device during use does not improve cutting performance and in extreme cases will cause irreversible damage to the device.
|